AC WITHSTAND VOLTAGE |
Output Rating | 5KVAC / 50mA | 5KVAC / 100mA |
Range | Range | Resolution | Accuracy |
Output Voltage ACV | 0 - 5000 | 1 | ± (2% of setting + 5V) |
Output Frequency | 50Hz / 60Hz ±0.1% User Selection |
Output Waveform | Sine Wave Crest Factor = 1.3 - 1.5 |
Output Regulation | ± (1% of output + 5V) From no load to full load and Low Line to High Line (combined regulation) |
SETTINGS |
HI and LO-Limit (Total) current mA | 0.000 - 9.999 | 0.001 | ± (2% of setting + 2 counts) |
10.00 - 50.00 (for ESA-140) | 0.01 |
10.00 - 100.00 (for ESA-150) |
HI and LO-Limit (Real) current mA | 0.000 - 9.999 | 0.01 | ± (3% of setting + 50µA) |
10.00 - 50.00 (for ESA-140) | 0.1 |
10.00 - 99.99 (for ESA-150) |
Ramp - Up Timer second | 0.1 - 999.9 | 0.1 | ± (0.1% of setting + 0.05 sec) |
Ramp - Down Timer second | 0.0 - 999.9 |
Dwell Timer second | 0 0.4 - 999.9 (0 = continuous) |
Ground Continuity | Current: DC 0.1 A ± 0.01A fixed Max. Ground Resistance: 1.0 Ω ± 0.1Ω |
Current Offset | 0.000 - 50.00mA (PF=0.00 - 1.00)* (Total current + current offset ≤50mA) for ESA-140 0.000 - 99.99mA (PF=0.00 - 1.00)* (Total current + current offset ≤100mA) for ESA-150 |
Arc Detection | 1 - 9 ranges (9 is the most sensitivity) |
DC WITHSTAND VOLTAGE |
Output Voltage DCV | 0 - 6000 | 1 | ± (2% of setting + 5V) |
DC Output Ripple | < 4% (6KV/20mA at Resistive Load) |
SETTINGS |
HI and LO - Limit current μA | 0.0 - 999.9 | 0.1 | ± (2% of setting + 2 counts) |
1000 - 20000 | 1 |
Ramp - Up Timer second | 0.4 - 999.9 | 0.1 | ± (0.1% of setting + 0.05 sec) |
Ramp - Down Timer second | 0.0 1.0 - 999.9 |
Dwell Timer second | 0 0.3 - 999.9 (0 = continuous) |
Ramp - HI current | > 20 mA peak maximum ON / OFF User Selection |
Charge - LO current | 0.0 - 350.0 µA DC or Auto Set |
Discharge Time | < 10mSec |
Maximum Capacitive Load DC Mode | 1µF < 1KV 0.08µF < 4KV |
0.75µF < 2KV 0.04µF < 5KV |
0.5µF < 3KV |
Ground Continuity | Current: DC 0.1 A ± 0.01A fixed Max. Ground Resistance: 1.0 Ω ± 0.1Ω |
Arc Detection | 1 - 9 ranges (9 is the most sensitivity) |
INSULATION RESISTANCE |
Output Voltage DCV | 30 - 1000 | 1 | ± (2% of setting + 2 counts) |
Charging Current | Maximum > 20mApeak |
SETTINGS |
HI and LO - Limit resistance MΩ | 0.05 - 99.99 (HI - Limit: 0 = OFF) | 0.01 | 0.05 - 999.9 ± (2% of setting + 2 counts) |
100.0 - 999.9 | 0.1 | 1000 - 9999 ± (5% of setting + 2 counts) |
1000 - 50000 | 1 | 10000 - 50000 ± (15% of setting + 2 counts) |
Ramp - Up Timer second | 0.1 - 999.9 | 0.1 | ± (0.1% of setting + 0.05 sec) |
Ramp - Down Timer second | 0.0 1.0 - 999.9 |
Dwell Timer second | 0 0.5 - 999.9 (0 = continuous) |
Delay Timer second | 0.5 - 999.9 |
Charge - LO current μA | 0.000 - 3.500 or Auto Set |
GROUND BOND |
Output AC Current A | 1.00 - 40.00 | 0.01 | ± (2 % of setting + 2 counts) |
Output Voltage Vac | 3.00 - 8.00 | 0.01 | ± (2 % of setting + 3 counts) |
Output Frequency Hz | 50Hz / 60Hz ± 0.1% User Selection |
Output Regulation | ± (1% of output + 0.02A) Within maximum load limits and over input voltage range |
Maximum Loading | 1.00 - 10.00A / 0 - 600mΩ 10.01 - 30.00A / 0 - 200mΩ 30.01 - 40.00A / 0 - 150mΩ |
SETTINGS |
Lead Resistance Offset mΩ | 0 - 200 | 1 | ± (2% of setting + 2 counts ) |
HI and LO - Limit Resistance mΩ | 0 - 150 (30.01 - 40.00A) | 1 | 6.00 - 40A ± (2% of setting + 2 counts ) |
0 - 200 (10.01 - 30.00A) |
0 - 600 (6.00 - 10.00A) |
0 - 600 (1.00 - 5.99A) | 1.00 - 5.99A ± (3% of setting + 3 counts ) |
Dwell Timer second | 0 0.5 - 999.9 (0 = continuous) | 0.1 | ± (0.1% of setting + 0.05 sec) |
CONTINUITY TEST |
Output Current | 0.1A for 0 - 10.00Ω 0.01A for 10.1 - 100.0Ω 0.001A for 101 - 1000Ω 0.0001A for 1001 - 10000 0.1A is Max. |
SETTINGS |
Max and Min - Limit Resistance Ω | 0.00 - 10.00 | 0.01 | ± (1% of setting + 3 counts) |
10.1 - 100.0 | 0.1 |
101 - 1000 | 1 |
1001 - 10000 | 1 | ± (1% of setting + 10 counts) |
Dwell Timer second | 0.0 0.3 - 999.9 (0 = continuous) | 0.1 | ± (0.1% of setting + 0.05 sec) |
Resistance Offset Ω | 0.00 - 10.00 | 0.01 | ± (1 % of reading + 3 counts) |
MEASUREMENT |
| Range | Resolution | Accuracy |
Voltage KV (AC / DC) | 0.00 - 6.00 | 0.01 | ± (2 % of reading +1 count) |
Voltage Vdc (IR only) | 0 - 1000 | 1 | ± (2 % of reading + 2 counts) |
AC Current (Total) mA | 0.000 - 3.500 | 0.001 | ± (2% of reading + 2 counts) |
3.00 - 100.00 | 0.01 |
AC Current (Real) mA | 0.000 - 9.999 | 0.001 | ± (3% of reading + 50µA) All Ranges PF > 0.1 V > 250VAC |
10.00 - 99.99 | 0.01 |
DC Current μA | 0.0 - 350.0 | 0.1 | ± (2% of reading + 2 counts) |
DC Current mA | 0.300 - 3.500 | 0.001 |
3.00 - 20.00 | 0.01 |
AC Current A (GB) | 0.00 - 40.00 | 0.01 | ± (3 % of reading + 3 counts) |
Resistance MΩ (IR) | 30 - 499V | 500 - 1000V | | 30 - 499V 0.05 - 999.9 ± (7% of reading+2 counts) 500 - 1000V 0.05 - 999.9 ±(2% of reading+2 counts) 1000 - 9999 ±(5% of reading+2 counts) 10000 - 50000 ±(15% of reading+2 counts) |
0.050 - 1.999 | 0.050 - 9.999 | 0.001 |
2.00 - 19.99 | 10.00 - 99.99 | 0.01 |
20.0 - 199.9 | 100.0 - 999.9 | 0.1 |
200 - 50000 | 1000 - 50000 | 1 |
Resistance mΩ (GB) | 0 - 600 | 1 | 1.00 - 2.99 A ± (3 % of reading + 3 counts) 3.00 - 40.00 A ± (2 % of reading + 2 counts) |
Resistance Ω (Continuity) | 0.00 - 10.00 | 0.01 | ± (1 % of reading + 3 counts) |
10.1 - 100.0 | 0.1 |
101 - 1000 | 1 |
1001 - 10000 | 1 | ± (1 % of reading + 10 counts) |
GENERAL |
Input Voltage AC | 115/230Vac±15% Auto Range 50/60Hz ± 5% 5A/250Vac Slow-Blow for ESA-140 10A/250Vac Slow-Blow for ESA-150 |
PLC Remote Control | Input : Test Reset Interlock Recall File 1 through 3 Recall File 1 through 7(Option) Output: Pass Fail Test-in-Process |
Memory | 10000 steps allow the user create different memories and steps |
TFT LCD | 800 x 480 resolution digital TFT LCD 9 ranges contrast setting |
DualCHEK | 5kVac/25mA and 25Aac/150mΩ for ESA-140 5kVac/50mA and 30Aac/150mΩ for ESA-150 |
Safety | Built-in Smart GFI circuitGFI trip current 5.0mA max. HV shut down speed: <1mS (on 50/60Hz and test under 1000V) |
Hot Hipot Tests | To detect the line input voltage to produce a simultaneous sine wave of line power at hipot output |
MyMenu | This is a menu that would be customizable by the user allowing them to create a menu of their favorite most used functions and menu selection. |
Interface | Standard USB & RS232 PC Control Card Optional Ethernet GPIB (IEEE-488.2) Multi-function Interface card (USB-A / RS-485 / RS-232 / BAR Code PS / 2 type) |
Multinational language | The operate screen interface to provide selection different language including English / Traditional Chinese / Simplified Chinese |
Alarm Volume Setting | Range: 0 - 9 ;0=OFF 1 is softest volume 9 is loudest volume. |
Calibration | Software and adjustments are made through the front panel. Automatic Calibration alert function to signal operator when re-calibration is due. |
Environment | 0 - 40ºC 20 - 80%RH |
Dimensions/Net Weight | 430mm(W) × 133mm(H) × 500 mm(D) / 30Kg |
OPTION |
Opt.794 MATRIX SCANNER MODULE |
High Voltage Rating | 5KVAC/6KVDC |
High Current Rating | 40 A |
Number of HV Channel | 8 |
Number of HA Channel | 8 |
Point to Point Continuity | Used the Scanner to reach point to point continuity test and this function is standard feature when built in scanner |
Opt.767 RUN TEST MODULE |
DUT POWER |
AC Voltage | 0 - 277.0V Single phase unblance |
Current | 16A maximum continuous |
Power Rating | 4500W maximum |
Short Circuit Protection | 23Arms or Inrush Current 68Apeak Response time RMS < 3 sec ; PEAK < 10uS |
SETTINGS |
HI and LO Limit AC Voltage V | 30.0 - 277.0 | 0.1 | ± (1.5% of setting + 0.2 V) |
HI and LO Limit AC Current A | 0.00 - 16.00 | 0.01 | ± (2% of setting + 2 counts) |
HI and LO Limit AC Power W | 0 - 4500 | 1 | ± (5% of setting + 3 counts) |
HI and LO Limit Power Factor | 0.000 - 1.000 | 0.001 | ± (8% of setting + 2 counts) |
HI and LO Limit Leakage Current | 0.00 - 10.00 Hi - Limit 0 = OFF | 0.01 | ± (2% of setting + 2 counts) |
Delay Time second | 0.2 - 999.9 | 0.1 | ± (0.1% + 0.05 sec) |
Dwell Time second | 0 0.1 - 999.9 (0 = continuous) |
MEASUREMENT |
| Range | Resolution | Accuracy |
Voltage Vac | 0.0 - 277.0 | 0.1 | ± (1.5% of reading + 2counts) 30 - 277V |
Current Aac | 0.00 - 16.00 | 0.01 | ± (2% of reading + 2counts) |
Power Watts | 0 - 4500 | 1 | ± (5% of reading + 3 counts) |
Power Factor | 0.000 - 1.000 | 0.001 | ± (8% of reading + 2 counts) |
Leakage Current mA | 0.00 - 10.00 | 0.01 | ± (2% of reading + 2 counts) |
MD | Leakage current measuring resistor = 2kΩ ± 1% |
Opt.768 TOUCH CURRENT TEST MODULE |
DUT |
DUT Input Power Rating | 0-277VAC@ 16Aac max |
Current | 16A maximum continuous |
Short Circuit Protection | 23Arms or Inrush Current 68Apeak Response time RMS < 3 sec ; PEAK < 10uS |
SETTINGS |
Leakage current High/Low Limit (RMS) µA | Range | 0.0-999.9µA (0=OFF) | Resolution | 0.1µA |
1000~10000uA | 1µA |
Leakage current High/Low Limit (peak) µA | Range | 0.0-999.9µA (0=OFF) | Resolution | 0.1µA |
1000~10000uA | 1µA |
Dwell Time second | 0 0.7-999.9 (0=continous) | 0.1 | ± (0.1% + 0.05 sec) |
Delay Time second | 0.5-999.9 |
Compliance Standards | A. UL544 Non Patient UL484 IEC60598 UL1363UL923 UL471 UL867 UL697 |
B. UL544 Patient Care |
C. UL2601-1 IEC60601-1 EN60601-1 |
D. UL1563 |
E. IEC60990 Fig4 U2 IEC 60950-1 IEC60335-1 IEC60598-1 UL484 IEC60065 IEC61010IEC60065 |
F. IEC60990 Fig5 U3 IEC60598-1 |
G. Basic measuring element 1k ohm of frequency check |
MD A - G components | Accuracy: Resistance ± 1% Capacitance ± 5% |
MD Voltage Limit | Maximum 30V peak or 30VDC |
Probe settings | G-L Ph-Pl Ph-L (Use HV relay and HV terminal connector) |
Internal Leakage | 1. Internal Leakage current = 65uA 2. 277V applied to Ph max leakage current = 70uA. |
External MD | User can add one extra MD for his application. |
CURRENT MEASUREMENT | The leakage current is fitting range by leakage current Hi-limit setting value |
Frequency Range | DC 15Hz≤ F ≤1MHz |
Leakage Current Range (RMS) |
Auto Range | Range 1 - Range 6 | 0.0uA - 10.00mA | Resolution | 0.1uA / 1uA / 0.01mA |
Fixed Range > 6% of Range | Range 1 - Range 6 | 0.0uA - 10.00mA | Resolution | 0.1uA / 1uA / 0.01mA |
Fixed Range < 6% of Range | Range 2 - Range 6 | 0.0uA - 600uA | Resolution | 0.1uA / 1uA / 0.01mA |
Accuracy For Auto Range |
Range | Mode | Frequency | Basic Accuracy |
Range 1~5*1 | AC+DC | DC | ±2% of reading + 3 counts |
15Hz < f <100kHz | ±2% of reading + 3 counts |
100kHz < f < 1MHz | ±5% of reading (>10.0uA) |
AC only*2 | 15Hz < f <30Hz | ±3% of reading + 5 counts |
30Hz < f <100kHz | ±2% of reading + 3 counts |
100kHz < f < 1MHz | ±5% of reading (>10.0uA) |
DC only*3 | DC | ±2% of reading + 3 counts(>10.0uA) |
Range 6*1 | AC+DC | DC | ±5% of reading (>10.0uA) |
15Hz < f <100kHz |
AC only*2 | 15Hz < f <30Hz |
30Hz < f <100kHz |
DC only*3 | DC |
Accuracy For Fixed Range |
Range | Mode | Frequency | Basic Accuracy(>6% of Range) | Additional Error(<6% of Range) |
Range 1~5*1 | AC+DC | DC | ±2% of reading + 3 counts | add (2% of reading + 0.2% of range) |
15Hz < f <100kHz | ±2% of reading + 3 counts | add (2% of reading + 0.2% of range) |
100kHz < f < 1MHz | ±5% of reading (>10.0uA) | add (2% of reading + 0.5% of range) |
AC+DC | 15Hz < f <30Hz | ±3% of reading + 5 counts | add (2% of reading + 0.2% of range) |
30Hz < f <100kHz | ±2% of reading + 3 counts | add (2% of reading + 0.2% of range) |
100kHz < f < 1MHz | ±5% of reading (>10.0uA) | add (2% of reading + 0.5% of range) |
DC only*3 | DC | ±2% of reading + 3 counts(>10.0mV) | add (2% of reading + 0.2% of range) |
Range 6*1 | AC+DC | DC | ±5% of reading (>10.0mV) | add (2% of reading + 0.2% of range) |
15Hz < f <100kHz |
AC only*2 | 15Hz < f <30Hz |
30Hz < f <100kHz |
DC only*3 | DC |
Leakage Voltage Range (Peak) |
Auto Range | Range 1 - Range 6 | 0.0mV - 15.00V | Resolution | 0.1mV / 1mV / 0.01V |
Fixed Range > 6% of Range | Range 1 - Range 6 | 0.0mV - 15.00V | Resolution | 0.1mV / 1mV / 0.01V |
Fixed Range < 6% of Range | Range 2 - Range 6 | 0.0mV - 900mV | Resolution | 0.1mV / 1mV / 0.01V |
Accuracy For Auto Range |
Range | Mode | Frequency | Basic Accuracy |
Range 1~5*1 | AC+DC | DC | ±2% of reading + 2mV |
15Hz < f < 1MHz | ±10% of reading +2mV |
AC only*2 | 15Hz < f < 1MHz | ±10% of reading +2mV |
Range 6*1 | AC+DC | DC | ±2% of reading + 3 counts |
15Hz < f <100kHz | ±10% of reading +2 counts |
AC only*2 | 15Hz < f <100kHz | ±10% of reading +2 counts |
Accuracy For Fixed Range |
Range | Mode | Frequency | Basic Accuracy(>6% of Range) | Additional Error(<6% of Range) |
Range 1~5*1 | AC+DC | DC | ±2% of reading + 2mV | add (2% of reading + 0.2% of range) |
15Hz < f <100kHz | ±10% of reading + 2mV | add (2% of reading + 0.2% of range) |
100kHz < f < 1MHz | ±10% of reading +2mV | add (2% of reading + 0.5% of range) |
AC only*2 | 15Hz < f <100kHz | ±10% of reading + 2mV | add (2% of reading + 0.2% of range) |
100kHz < f < 1MHz | ±10% of reading 2mV | add (2% of reading + 0.5% of range) |
Range 6*1 | AC+DC | DC | ±2% of reading +3 counts | add (2% of reading + 0.2% of range) |
15Hz < f <100kHz | ±10% of reading +2 counts |
AC only*2 | 15Hz < f <100kHz | ±10% of reading +2 counts |
To exin with notes for leakage |
*1.If the final measured signal is >range 5 then the maximum composite signal that can be measure is 28 volts peak。 If the final measured signal is ≤ range 5 then the maximum composite signal that can be measure is 12 volts peak。 |
*2.AC cutoff frequency for High Pass Filter is 15Hz on AC only mode。 |
*3.AC cutoff frequency for Low Pass Filter is 15Hz on DC only mode。 |
Leakage Imax Range |
The specification is as same as leakage current(RMS) |
The specification is as same as leakage current(Peak) |
LINE VOLTAGE MEASUREMENT |
Range | 0.0-277.0 VAC |
Resolution | 0.1 V |
Accuracy | ± (1.5% of reading +0.2V) 30.0-277.0V |
GENERAL |
Continous Power Output selection (like Continuous Run) for both TCT and Run testing. | To create Continnuous parameter selection for both TCT and RUN testing.When continuous =ON the power is not shut down between steps .But in the TCT mode when the steps setting different line condition DUT output will momentary power off 25ms then power on. |
Opt.769 AC SOURCE 500VA |
OUTPUT |
Power | 630VA and 500W Maximum |
Voltage | 0 - 150.0V / 0 - 277.0V |
Current | 4.20A / 2.10A |